Título : A novel device for temporary cardiac stimulation using permanent pacemaker generators and active-fixation leads: the TEMPACE multicentre prospective study |
Autor : Vicente-Miralles, Raimundo  Rivas-Gándara, Nuria  Jordán Marchite, Pablo Adeliño, Raquel  Carmona, María Ángeles Díaz Barranco, Mónica Ibañez Criado, Alicia Ascensión Ruiz Nodar, Juan Miguel Ajo Ferrer, María  García-Fernández, Amaya  Bertomeu-González, Vicente  Martínez Martínez, Juan Gabriel |
Editor : European Society of Cardiology |
Departamento: Departamentos de la UMH::Medicina Clínica |
Fecha de publicación: 2025-08 |
URI : https://hdl.handle.net/11000/37867 |
Resumen :
To determine the safety and effectiveness of the KronoSafe® permanent pacemaker adapter for temporary cardiac pacing (TCP) with active-fixation leads (TCPAFL). Methods: This was a multicenter, prospective, descriptive clinical investigation involving a medical device (ClinicalTrials.gov Identifier: NCT05351658). Between January 2023 and December 2024, all consecutive patients who underwent implantation of a temporary active-fixation pacemaker were included. The devices were secured using the KronoSafe® adapter throughout the period of use. Complications during TCPAFL were recorded, and R-wave sensing, lead impedance, and pacing threshold were determined every 48 hours. Results: Thirty patients (10 with atrioventricular block, 9 TAVI, 5 alcohol septal ablations, 2 slow atrial fibrillations, 2 bradycardia-tachycardia syndrome, and 2 pacemaker infections) who required TCP were included. Two centers participated in the study. The mean duration of TCP was 7.8 days (maximum 22 days), with 79.2% of the time spent in the cardiology ward. One complication (3%) was recorded, due to accidental traction of the system associated with an episode of agitation. Conclusions: TCPAFL using the KronoSafe® adapter is safe and effective, and its use can be extended outside the Intensive Care Unit.
|
Palabras clave/Materias: external permanent pacemaker temporary pacemaker temporaryt pacemaker with active-fixation leads temporary-permanent pacemaker |
Tipo de documento : info:eu-repo/semantics/article |
Derechos de acceso: info:eu-repo/semantics/openAccess Attribution-NonCommercial-NoDerivatives 4.0 Internacional |
DOI : 10.1093/ehjacc/zuaf111 |
Publicado en: Eur Heart J Acute Cardiovasc Care . 2025 Aug 28:zuaf111 |
Aparece en las colecciones: Artículos Medicina Clínica
|